CT Scan

                                                        MACHINE FEATURES

Constructor

Fraunhofer IKTS

High voltage X-ray tube

Open microfocus x-ray tube max operating voltage  300 kV

X-Ray Detector

2048 x 2048 pixels, size 400 x 400 mm,

resolution 200 μm

Manipulator stage

 4-axis manipulation system allowing:

-      sample manipulation xyz;

-      sample rotation full range 0-360° and continuous rotation

X-ray image resolution

5 μm (under most favorable condictions)

Type of investigable materials 

- Metals (iron-based and non-iron-based alloys)

- Crystalline ceramics

- Glasses

- Polymers

- Composites

- Multimaterials devices with different densities

  (from low-density polymers to high-density metals)

 - Multilayer materials comprising joining interfaces

 - Bulk and porous materials

Sample size

Max. 500 mm x 500 mm x 500 mm (L x W x H)

Sample mass 

Max 100 kg

Mechanical test

Tensile, compression and bending test up to 10 kN of load during x-ray inspection