Vacuum Furnace

                                                      FURNACE FEATURES

Constructor

XERION BERLIN LABORATORIES

Model

XVAC Series

Max. temperature in the working zone

1700 °C

Max. heating rate

10 K/min

Working atmosphere

Vacuum/Inert gas

Temperture measurement

- 2 TC Type C for temperature control

- 2 TC Type C for overtemperature 

- 1 Pyrometer

- ± 5 K homenity of temperature

- ± 2 K accuracy of temperature

Working space specification

220 mm x 220 mm x 220 mm

Vacuum system

- Turbomolecular pump and Rotary vane pump

- Final vacuum 5 x 10-6 mbar in empty and cold furnace

Gas system

- Argon

- Mass flow controller 1-10 l/min

Control specifications

- Eurotherm 2404

- Up to 20 programmes with 16 segments

Chiller specifications

- Cooling medium water

- Temperature/Pressure cooling medium 25 °C/2-4 bar

Camera specifications

- Pyrometer with integrated camera

- Pyrometer temperature range 200 °C - 1700 °C

- Spectral range of pyrometer 1,5 μm to 1,8 μm